聚焦超声换能器电阻抗谱实时在线监测系统
引用本文:肖建,黄文军,刘雅璐,罗曼,王杰,宋丹. 聚焦超声换能器电阻抗谱实时在线监测系统[J]. 压电与声光, 2024, 46(3): 420-428
作者姓名:肖建  黄文军  刘雅璐  罗曼  王杰  宋丹
作者单位:重庆医科大学 生物医学工程学院、超声医学工程国家重点实验室,重庆 400016 ;重庆医科大学 重庆市生物医学工程学重点实验室,重庆 400016;超声医疗国家工程研究中心,重庆 401121;重庆医科大学 生物医学工程学院、超声医学工程国家重点实验室,重庆 400016 ;超声医疗国家工程研究中心,重庆 401121
基金项目:国家自然科学基金委员会青年基金(12204079);重庆市自然科学基金面上项目(CSTB2023NSCQ-MSX0532)
摘    要:高强度聚焦超声(HIFU)换能器作为聚焦超声消融手术(FUAS)系统的核心部件,其电阻抗谱是确定FUAS系统工作参数的重要参考依据,故对其进行实时在线监测具有重要意义。该文采用脉冲响应时域测量方法,将探测脉冲信号与HIFU 换能器的工作时序相融合,并基于LabVIEW 平台构建了一套HIFU 换能器电阻抗谱实时在线监测系统,实现了对HIFU 换能器电阻抗幅值与相位的高重复性及准确性的实时测量。通过对长时工作在大功率、高占空比条件下HIFU 换能器电阻抗谱的实时在线监测,表明该监测系统可准确反应HIFU 换能器的特征频率漂移、阻抗幅值和相位的非线性变化。这为HIFU 换能器电阻抗谱实时在线监测及驱动参数优化调节提供了技术方案,对HIFU 换能器的设计优化和能量损耗机制研究具有重要的应用价值。

关 键 词:高强度聚焦超声(HIFU)换能器;电阻抗谱;脉冲响应;实时在线监测
收稿时间:2024-01-16

Real-Time Online Monitoring System for Electric Impedance Spectra of Focused Ultrasound Transducers
XIAO Jian,HUANG Wenjun,LIU Yalu,LUO Man,WANG Jie,SONG Dan. Real-Time Online Monitoring System for Electric Impedance Spectra of Focused Ultrasound Transducers[J]. Piezoelectrics & Acoustooptics, 2024, 46(3): 420-428
Authors:XIAO Jian  HUANG Wenjun  LIU Yalu  LUO Man  WANG Jie  SONG Dan
Affiliation:State Key Laboratory of Ultrasound in Medicine and Engineering, College of Biomedical Engineering,Chongqing Medical University, Chongqing 400016 , China ;Key Laboratory of Biomedical Engineering, Chongqing Medical University, Chongqing 400016 , China;National Engineering Research Center of Ultrasonic Medicine, Chongqing 401121 , China;State Key Laboratory of Ultrasound in Medicine and Engineering, College of Biomedical Engineering,Chongqing Medical University, Chongqing 400016 , China ;National Engineering Research Center of Ultrasonic Medicine, Chongqing 401121 , China
Abstract:

High-intensity focused ultrasound (HIFU) transducers are a core component of focused ultrasound

ablation surgery (FUAS) systems, whose electric impedance spectrum is an important index to determine working

parameters that significantly affect FUAS systems performance.Real-time online monitoring of this spectrum is significantly

beneficial.In this study, based on pulse response time-domain measurement, by merging the detected

pulse signal serial into the FUAS operational electric driving signal sequence of the HIFU transducer, a real-time online

monitoring system for the electric impedance spectrum of the HIFU transducer was constructed on the Lab-

VIEW platform.Highly reproducible accurate measurements towards the magnitude and phase spectrum of the HIFU

transducer electric impedance were achieved.Subsequently, a monitoring system was applied to determine dynamic

changes in the electric impedance of the HIFU transducer under extreme working conditions.These include

long-time service, high electrical driving power and duty cycle.The results demonstrate that the proposed system accurately

monitors the HIFU transducer eigen frequencies drift, alongside nonlinear changes in impedance magnitude

and phase.This paper presents a technical solution for real-time online monitoring of HIFU transducer impedance

during FUAS, and provides feedback for drive parameter adjustment, which has practical application value for HIFU

transducer design optimization and energy loss mechanisms.

Keywords:high-intensity focused ultrasound (HIFU) transducer; electric impedance spectrum; pulse response;real-time online monitoring