Authors:Fu?Xin?  author-information"  >  author-information__contact u-icon-before"  >  mailto:xfu@zju.edu.cn"   title="  xfu@zju.edu.cn"   itemprop="  email"   data-track="  click"   data-track-action="  Email author"   data-track-label="  "  >Email author,Xie?Haibo,Yang?Huayong,Jia?Zhijian,Fang?Qun
Affiliation:1. State Key Laboratory of Fluid Power Transmission and Control,Zhejiang University,Hangzhou 310027,China
2. Institute of Microanalytical Systems,Zhejiang University,Hangzhou 310028,China
Abstract:A new structure and working principle of the chip integrated with micro valveless pump for capillary electrophoresis was proposed in this paper. The micro valveless pump with plane structure has advantages of simple structure, and the process technology is compatible with existing micro chips for capillary electrophoresis. Based upon the mathematical model, simulation study of micro pump was carried out to investigate the influence of structural parameters on flow characteristics, and the performance of the integrated micro pump was also tested with different control parameters. The simulation results agree with the experimental results. Three samples, which are amino acid, fluorescein and buffer solution, have been examined with this chip. The results of the primary experiments showed that the micro valveless pump was promising in the integration and automatization of miniature integrated fluidic systems.