A sensitive EPR dosimetry system based on sulfamic acid
Authors:A Maghraby
Affiliation:National Institute of Standards (NIS), Radiation Dosimetry Department, Tersa St., P.O. Box 136, 12211 Giza, Egypt
Abstract:There is a need for a sensitive dosimeter using electron paramagnetic resonance spectroscopy (EPR) for use in medical applications, since with this method non-destructive read-out and dose archival could be achieved. Sulfamic acid as a possible detector substance was investigated before and after irradiation, its EPR spectra were recorded and analyzed, some dosimetric characteristics were studied: microwave power saturation behavior, the effect of modulation amplitude on peak-to-peak signal intensity and the line width. Energy-dependence parameters were compared to soft tissue and alanine, and the response to ionizing radiation was studied, also the decay behavior along 133 days after irradiation is presented. It is found that sulfamic acid possesses high-sensitivity and reasonable signal stability which may make it useful as a sensitive dosimeter for medical applications.
Keywords:28  41  Te  42  88  +h  61  80  &minus  x  61  80  Lj  61  82  &minus  d  87  50  Gi  87  53  &minus  j  87  66  &minus  a  33  35  +r  76  30  &minus  v