首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of the force required to move a neurosurgical probethrough in vivo human brain tissue
Authors:Howard  MA  III Abkes  BA Ollendieck  MC Noh  MD Ritter  C Gillies  GT
Affiliation:Division of Neurosurgery, University of Iowa Hospitals and Clinics, Iowa City 52242, USA.
Abstract:The advent of high-precision magnetic and robotic computer-controlled neurosurgery systems makes it necessary to determine the range of forces that will be encountered by the probes of such devices as they are guided through the brain tissues to intraparenchymal targets. We have measured the penetration forces on 2.5-mm spheres and the drag forces on 3.0-mm ventricular shunt catheters advanced 2.0-3.5 cm deep into in vivo human brain tissues (in patients about to have those tissues resected during epilepsy surgery) at rates of approximately 0.33 mm s-1. Penetration forces of (8 +/- 2) grams were found for the spherical probe once it passed 0.5 cm below the cortical surface, and frictional drags of (2.8 +/- 0.3) grams cm-1 were exerted on the catheters. The variable nature of these forces is discussed and the results are compared with earlier studies on experimental animal tissues and brain phantom gelatins. The implications of these results for magnetic and robotic surgery systems are considered.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号