首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of high‐resolution and mobile display systems for digital radiology in dark and bright environments using human and computational observers
Authors:Rebecca Vogel  Anindita Saha  Kish Chakrabarti  Aldo Badano
Abstract:Abstract— As digital display systems replace film traditionally used for reading radiographic images, resource‐intensive acceptance testing must be performed to ensure that quality meets and maintains desired specifications. If machine observers can replace human readers, whose performances are highly variable, the results will be more consistent and less costly. To be effective, however, the automated observers must track human performance. An approach for a model observer, validated with human readers, for the evaluation of the visibility of low‐contrast small targets in high‐resolution and mobile displays under different ambient illumination, will be described. The displays were tested using CDMAM‐like digital phantoms containing disks of varying diameters and contrasts on a flat background. For this task, we find the best indicator of display performance to be the display's ability to represent small luminance contrast, not resolution or pixel size. The results confirm that high‐resolution systems perform better under low illumination while illuminance has a minor impact on the mobile‐display performance. Finally, the results show that the machine observer tracks the performance of human readers. Machine observers with proper validation can replace humans in the acceptance testing procedures, saving the testers both time and money.