首页 | 本学科首页   官方微博 | 高级检索  
     

基于极大倾角光纤光栅的降钙素原免疫传感器
引用本文:赵明富,徐杨非,罗彬彬,石胜辉,蒋棚俊,鲁姣,唐平,吴胜昔. 基于极大倾角光纤光栅的降钙素原免疫传感器[J]. 半导体光电, 2018, 39(1): 10-14,20. DOI: 10.16818/j.issn1001-5868.2018.01.003
作者姓名:赵明富  徐杨非  罗彬彬  石胜辉  蒋棚俊  鲁姣  唐平  吴胜昔
作者单位:重庆理工大学光纤传感与光电检测重庆市重点实验室,重庆,400054;重庆理工大学光纤传感与光电检测重庆市重点实验室,重庆400054;重庆理工大学电子科技大学光电信息学院,成都610054;重庆理工大学药学与生物工程学院,重庆,400054
基金项目:国家自然科学基金项目,重庆市科委基础与前沿研究项目,中国博士后基金项目,重庆市教委基础/应用基础研究项目,重庆理工大学研究生创新基金重点项目
摘    要:降钙素原(Procalcitonin,PCT)对于反映患者炎症甚至败血症患病程度具有重要意义。为实现低浓度PCT的免标记检测,利用极大倾角光纤光栅(Excessively Tilted Fiber Grating,Ex-TFG)的低温度交叉敏感性和高折射率灵敏度的优点,在光纤的表面修饰金黄色葡萄球菌(Staphylococcus Aureus Protein A,SPA),再将PCT单克隆抗体固定在SPA分子层,从而构成对PCT具有特异性检测功能的免疫传感器。实验结果表明:降钙素原含量为0.5ng/mL时,谐振波长红移约13pm;PCT的浓度为200ng/mL时,Ex-TFG的红移量基达约125pm;浓度为1 000ng/mL时,红移量趋于饱和,最大红移量约150pm。通过对实验数据拟合,发现Ex-TFG谐振波长的红移量与PCT浓度之间的关系满足Langmuir模型,相关性系数为0.98。该方法为低浓度PCT的检测提供了一种免标记、快速的检测方法。

关 键 词:极大倾角光纤光栅  免疫传感器  降钙素原  免标记
收稿时间:2017-08-22

Study on Procalcitonin Immunosensor Based on Excessively Tilted Fiber Grating
ZHAO Mingfu,XU Yangfei,LUO Binbin,SHI Shenghui,JIANG Pengjun,LU Jiao,TANG Ping,WU Shengxi. Study on Procalcitonin Immunosensor Based on Excessively Tilted Fiber Grating[J]. Semiconductor Optoelectronics, 2018, 39(1): 10-14,20. DOI: 10.16818/j.issn1001-5868.2018.01.003
Authors:ZHAO Mingfu  XU Yangfei  LUO Binbin  SHI Shenghui  JIANG Pengjun  LU Jiao  TANG Ping  WU Shengxi
Abstract:Procalcitonin (PCT) plays an important role as the factor to reflect patients' inflammation or sepsis degree.In order to reach a low concentration detection of PCT with label free method,proposed was a type of PCT immunosensor based on excessively tilted fiber grating (Ex-TFG) which is insensitive to temputere but sensitive to refractive index.Staphylococcus aureus Protein A (SPA) was used to modify the surface of Ex-TFG,then highly purified PCT monoclonal antibodies (MAbs) were immobilized on Ex-TFG's surface through SPA molecules,so as to form the immunosensor which is specific to PCT.Experimental results show that the wavelength shift is about 13 pm when PCT concentration is 0.5 ng/mL,and it becomes to be 125 pm with the concentration of PCT increasing to 200 ng/mL.What's more,the immunosensor becomes saturated at ~ 1 000 ng/mL,and the corresponding wavelength shift is about 150 pm.Finally,by fitting the experimental data,it demonstrats that the fitting line meets the Langmuir model function with a corelated coefficient of 0.98.
Keywords:excessively tilted fiber grating(Ex-TFG)  immunosensor  procalcitonin(PCT)  label free
本文献已被 维普 万方数据 等数据库收录!
点击此处可从《半导体光电》浏览原始摘要信息
点击此处可从《半导体光电》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号