首页 | 本学科首页   官方微博 | 高级检索  
     


Topographic and material contrast in low-voltage scanning electron microscopy
Authors:J Hejna
Abstract:Two scintillation backscattered electron (BSE) detectors with a high voltage applied to scintillators were built and tested in a field emission scanning electron microscope (SEM) at low primary beam energies. One detector collects BSE emitted at low take-off angles, the second at high takeoff angles. The low take-off detector gives good topographic tilt contrast, stronger than in the case of the secondary electron (SE) detection and less sensitive to the presence of contamination layers on the surface. The high take-off detector is less sensitive to the topography and can be used for detection of material contrast, but the contrast becomes equivocal at the beam energy of 1 keV or lower.
Keywords:low-voltage scanning electron microscope  backscattered electrons  detectors  material contrast  topographic contrast