Abstract: | An optical spectrometer for the visible range has been developed for the confocal scanning laser microscope (CSLM) Phoibos 1000. The spectrometer records information from a single point or a user-defined region within the microscope specimen. A prism disperses the spectral components of the recorded light over a linear CCD photodiode array with 256 elements. A regulated cooling unit cools the diode array, thereby reducing the detector dark current to a level, which allows integration times of up to 60 s. The spectral resolving power, λ/Δλ, ranges from 400 at λ = 375 nm to 100 at λ = 700 nm. Since the entrance aperture of the spectrometer has the same diameter as the detector aperture of the CSLM, the three-dimensional spatial resolution for spectrometer readings is equivalent to that of conventional confocal scanning, that is, down to 0.2 μm lateral and 0.8 μm axial resolution with an N.A.=1.3 objective. |