首页 | 本学科首页   官方微博 | 高级检索  
     


Foreword
Authors:W C Martin
Affiliation:Group Leader, Atomic Spectroscopy, Atomic Physics Division, National Institute of Standards and Technology
Abstract:The spectrum of a platinum hollow-cathode lamp containing neon carrier gas was recorded photographically and photoelectrically with a 10.7 m normal-incidence vacuum spectrograph. Wavelengths and intensities were determined for about 5600 lines in the region 1130–4330 Å. An atlas of the spectrum is given, with the spectral lines marked and their intensities, wavelengths, and classifications listed. Lines of impurity species are also identified. The uncertainty of the photographically measured wavelengths is estimated to be ± 0.0020 Å. The uncertainty of lines measured in the photoelectric scans is 0.01 Å for wavelengths shorter than 2030 Å and 0.02 Å for longer wavelengths. Ritz-type wavelengths are given for many of the classified lines of Pt II with uncertainties varying from ±0.0004 to ± 0.0025 Å. The uncertainty of the relative intensities is estimated to be about 20%.
Keywords:hollow-cathode lamp  neon  platinum  spectral atlas  spectrum  wavelength
设为首页 | 免责声明 | 关于勤云 | 加入收藏