U.K. National Radiological Protection Board Radon Calibration Procedures |
| |
Authors: | K D Cliff |
| |
Affiliation: | National Radiological Protection Board, Chilton, Didcot, Oxon, OX11 ORQ, U.K. |
| |
Abstract: | A procedure for the calibration of instruments for the detection of 222Rn in air is described. The method is based on the alpha-spectrometric determination of the concentration in air of 218Po in the calibration chamber. The calibration chamber is described, together with the method of maintaining a high aerosol concentration. The 218Po concentration at steady state in the chamber is found to be 98% of the 222Rn concentration typically. An assessment of the sources of uncertainty in the method presented indicate that the 222Rn concentration in the chamber can be determined with an overall uncertainty of about 7% at the 95% confidence level. |
| |
Keywords: | aerosol generation alphaspectrometry calibration chamber radon daughter measurement radon gas measurement statistical uncertainty |
|
|