An integrated semicompliant balloon ultrasound catheter for quantitative feedback and image guidance during stent deployment
Authors:Choi Charles D  Savage James  Stephens Douglas N  O'Donnell Matthew
Affiliation:Department of Biomedical Engineering, University of Michigan, Ann Arbor 48109-209, USA. cdchoi@umich.edu
Abstract:An integrated balloon ultrasound catheter prototype was designed to image from inside the balloon for real-time guidance during stent deployment. It was fabricated using a semicompliant balloon material (polyethylene) and a 20 MHz, 64-element circumferential ultrasound array. A commercial stent, nominally 4.4 mm in diameter and 12 mm in length, was used for a phantom study and placed along the length of the integrated balloon ultrasound catheter. A rubber phantom was created with an elastic modulus of 175 kPa with a 4.36 mm diameter lumen. Real-time balloon pressure measurements were recorded using a digital pressure sensor, and real-time radio-frequency (RF) data were captured as the balloon was inflated. The slope of the area-pressure ratio (APR) was compared to a reference measure of the balloon and stent expanded in water to determine a measure for optimal stent deployment. The results clearly indicate stent deployment at 11.1 atm using this metric. The APR slope could serve as quantitative feedback parameter for guiding stent deployment to reduce arterial injury and subsequent restenosis. After the stent deployment experiment, RF data were captured as the balloon catheter was moved along the length of the stent in pullback mode to confirm successful stent deployment. Ultimately, an integrated balloon ultrasound catheter could serve as a single catheter intervention device by providing real-time intravascular ultrasound (IVUS) imaging and quantitative feedback during stent deployment.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号