化学发光法测定超痕量锇
引用本文:宋正华,刘秦茵.化学发光法测定超痕量锇[J].工业催化,1996,4(4):57-61.
作者姓名:宋正华  刘秦茵
作者单位:西北大学分析测试研究中心,西安 710069; 陕西省气象局,西安 710015
摘    要:本文提出一种新的高灵敏度测定锇的化学发光方法, 检测下限可达1.8×10-14 g/mL,对1.0×10-13 g/mL锇(Ⅷ)进行13次平行测定, 相对标准偏差为4.2%。发光强度与锇浓度在3.3×10-13 g/mL~1.3×10-9 g/mL范围内呈线性关系。本法操作简单, 重现性好, 灵敏度高, 是侧定超痕量锇的一种有效方法。

关 键 词:化学发光  测定    催化  

Determination of Trace Osmium by Chemiluminescence Analysis
Song Zhenghua,Su Wenjie,Liu Qinyin.Determination of Trace Osmium by Chemiluminescence Analysis[J].Industrial Catalysis,1996,4(4):57-61.
Authors:Song Zhenghua  Su Wenjie  Liu Qinyin
Affiliation:Instrumental Anal & Res Centre, Northwest Univ, Xi′an, 710069;Shaanxi Provincial Meteorological Bureau, Xi′an, 710015
Abstract:A new and highly-sensitive method for determination of trace Os(Ⅷ) was developed on the basis of Os(Ⅷ) catalysted reaction of Luminol and O2 which generates chemiluminescence. The detection limit of this method is as low as 1.8×10-14 g/mL for Os(Ⅷ) and the relative standard deviation (RSD) for determination of 1.0×10-13 g/mL Os(Ⅷ)(n=13) is 4.2%. These is a linear relationship between chemiluminescence and Os(Ⅷ) concentration in the range of 3.3×10-13 g/mL to 1.3×10-9 g/mL. The method proved to be simple and effective for determination of trace osmium, with good reproduceability and high sensitivity.
Keywords:determination  trace osnium  catalytic reaction  chemiluminescence  